THE ENGINEERING RECORD, FILE 02 OF 19

Actuators

100 Hz150 Hz200 Hz250 Hz300 Hz350 Hz400 HzPacinian channel, most sensitive 200 to 300 Hz, peak near 250LRA 175 to 235 Hzrise to rated amplitude37 msstop, with DRV2605L braking35 to 50 msnatural stop, no braking177 msReference part Precision Microdrives C10-100; second source Vybronics VG1040003D, about 10 ms to half amplitude. Measured at the skin through the band liner before any perceptual claim.
Why a linear resonant actuator, and where it parks its energy. The Pacinian channel is most sensitive from 200 to 300 Hz with its peak near 250 Hz. A coin LRA resonates near 175 to 235 Hz, at the lower edge of that band, 175 Hz for the reference part, driven closed loop by the DRV2605L. The reference part rises to rated amplitude in about 37 milliseconds and stops in about 35 to 50 with braking, against 177 unbraked. An ERM couples frequency to amplitude, so stronger also means a different pitch, which blurs the second law, and it rises and brakes slowly, which blurs the third. Source: docs/02.

Decision: coin linear resonant actuator (LRA), 10 mm diameter, about 3.5 mm tall, resonant near 175 to 235 Hz, driven closed loop by the DRV2605L. Reference part Precision Microdrives C10-100 (175 Hz, about 1.5 G, 37 ms to rated amplitude, 177 ms natural stop, about 35 to 50 ms with braking). Second source Vybronics VG1040003D (about 2.0 G, about 10 ms to half amplitude).

Finalized order, 2026-09: the Vybronics VG0832022D 8 mm coin LRA is the default motor across the bands, with the VG1040003D 10 mm reserved for the waist and back zones that want more delivered acceleration. The C10-100 stays as the reference part the psychophysics numbers were modeled on.

Why LRA: the Pacinian channel is most sensitive from 200 to 300 Hz with the peak near 250 Hz, and an LRA parks all its energy there. ERM motors couple frequency and amplitude, so "stronger" also means "different pitch," which blurs Law 2, and they rise and brake slowly, which blurs Law 3. Voice coils and piezos are cleaner but too costly or need high voltage on the skin.

Alternatives rejected: 10 mm pancake ERM (cheap, no driver needed, poor control); voice coil tactors (lab grade, expensive); piezo (tens of volts on skin).

Measurement plan: delivered acceleration at the skin through the actual band liner, rise time to 50 percent and to rated, brake time with DRV2605L braking on. These replace the modeled motor stage in the latency budget.

Source: docs/02-actuators.md, the engineering record. Every number on this page is read from that file. If a fact is not in the record, it is measured on the bench, or it stays off this site.